![Westfalia 82 04 51 Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/westfalia/82-04-51/82-04-51_instruction-manual_963341061.webp)
56
Operation
Measuring Duty via Current Clamp
1. First remove the measuring cables before measuring via the
current clamp!
2. Set the rotary function switch (13) to the
A
range position
respectively to the
A~
position.
3. Press the current clamp trigger (4) to open the clamp. Fully
enclose the conductor to be measured with the current clamp.
4. Press the frequency/duty switch key (11) as many times until
the icon “%” appears on the right side of the display in order to
activate the duty cycle measuring mode.
5. Read the duty cycle value from the display.
Notes:
Do not exceed the maximum input current of 600 A rms AC.
Do not enclose more than one conductor in the clamp during
measuring. Incorrect measurement values may be obtained.
If the duty cycle is less than 10%, the symbol “
UL
” appears on
the display. If the duty cycle is higher than 94.9%, the overload
indicator “
OL
” appears on the display.
A frequency range of 10 Hz to 1 kHz can be displayed. It is
possible to test frequencies which are higher than 1 kHz, but
the accuracy of the test result cannot be ensured.
Duty Cycle Measurements via Clamp - Accuracy
Range Resolution
Accuracy
10 – 95%
0.1 %
± 3%
Measurement range: 10 – 1 kHz
Input current range:
≥
1 A rms AC (higher input current at higher
frequencies)
Max. input current: 600 A rms AC
Содержание 82 04 51
Страница 1: ...Westfalia Bedienungsanleitung Nr 109889...
Страница 4: ...III bersicht Overview 16...
Страница 39: ...34 Notizen...
Страница 69: ...64 Notes...