2.4 Setting the Sample Volume
1.
To set or adjust the sample volume, locate the external adjustment assembly, and then slide the dustcover down to
expose the external adjustment chamber (
).
Setting the sample volume is accomplished by:
• determining where the adjustment pin will need to be placed to attain the desired sample volume,
• rotating the upper external adjustment ring in relation to the adjustment pin setting to attain the desired sample
volume, and
• installing the adjustment pin to the predetermined slot.
The sample volume is determined by the distance between the upper external adjustment ring and the adjustment pin.
Once the desired volume is set, the sample volume can be changed at any time externally without removing the inFlow™ ACE
from the pipeline.
Figure 7: Upper Housing Diagram
2.
Determine to which slot the adjustment pin will be installed after the external adjustment rings have been set (
16
IOM-161
MODEL: INFLOW™ ACE WITH AI CONTROL™
REV: D
13839 West Bellfort Street, Sugar Land, TX 77498
welker.com
Service Department 281.491.2331