3. Remove the two M3x5 screws from the ECG board using a #1 Phillips bit.
4. Push back slightly on the two latches to remove the ECG board from the bottom
assembly.
Reassembly note
•
Use a #1 Phillips bit and a Torque driver calibrated to 5.0 in-lb ± 0.5 in-lb to secure
the ECG board screws.
Remove the ECG ferrite
•
Separate the top housing assembly from the bottom housing assembly as described.
Gently open the two side-catches to remove the ECG ferrite and the ECG board
harness from the printer top assembly.
Note
Set the harness cable aside for reuse during reassembly.
88 Disassembly, repair, and reassembly
CP 150 12-lead resting electrocardiograph
Содержание CP 150
Страница 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Страница 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Страница 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Страница 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Страница 27: ...Home screen Home screen The Home screen includes the following areas 23...
Страница 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Страница 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Страница 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Страница 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Страница 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Страница 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Страница 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Страница 133: ......