4. Specify settings.
Setting
Action/Description
DHCP
Select or deselect DHCP. Select DCHP to auto-connect
through Ethernet. Deselect DHCP to manfully enter
settings.
Network IP address
Touch
and enter the IP address to manually set up
the device for Ethernet communications.
Subnet mask
Touch
and enter the Subnet mask.
Gateway
IP address that routes packets to other networks. Touch
and enter the Gateway address.
DNS Server 1
IP address of a server running DNS services for locating
computers and services through user-friendly names.
Touch
and enter the DNS Server address.
DNS Server 2
Touch
and enter the DNS Server 2 address.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home screen, touch
Exit
.
Specify radio settings
Note
Your model might not contain all of these features.
Note
Radio features are enabled through hardware detection.
1. Access the Advanced Settings.
a. Touch the
Settings
tab.
b. Touch the
Advanced
tab.
c. Enter
6345
as the
Advanced settings code
.
d. Touch
OK
.
The General tab appears at the bottom of the screen and the Regional tab appears at
the top of the screen.
2. Touch the
Network
tab.
The Status tab appears at the top of the screen and the vertical Ethernet and Radio
tabs appear.
3. Touch the
Radio
tab to access WiFi and Radio settings.
4. Specify Radio Configuration settings.
Setting
Action/Description
Enable radio
Enable the radio for device communications. When
disabled, the radio is not available.
ESSID
Identification name of an 80211 wireless network. Touch
and enter the service set identifier (SSID- the
name of the access point). Enter a maximum of 32
characters.
32 Settings
CP 150 12-lead resting electrocardiograph
Содержание CP 150
Страница 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Страница 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Страница 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Страница 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Страница 27: ...Home screen Home screen The Home screen includes the following areas 23...
Страница 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Страница 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Страница 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Страница 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Страница 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Страница 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Страница 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Страница 133: ......