Perform the full-functional tests
Complete the full-functional CP 150 tests to verify device functionality.
Note
The functional verification tests help to confirm the proper operation of the
device and its options. The tests may also be useful as a diagnostic tool to
help isolate a malfunction.
The following off-the-shelf tools are required to complete the functional verification tests:
•
Standard ECG Patient Simulator (example: Model 430B, Medi Cal Instrument, Inc. or
equivalent)
•
USB Cable (Type A connector to mini-B connector)
•
Standard USB Thumb drive (256M to 64Gb, recommend 2Gb)
•
RJ45 Ethernet cable (less than 3 M in length)
1. From the
Home
screen, touch the
Settings
tab.
2. Touch the
Advanced
tab.
3. Enter
6345
as the access code and touch
OK
.
The General tab appears at the bottom of the screen and the Regional tab appears at
the top of the screen.
4. Touch the
Service
tab.
5. Touch the
Self-tests
tab.
Test the hardware components.
6. Touch
CP 150
to perform the 9 hardware tests.
50 Service
CP 150 12-lead resting electrocardiograph
Содержание CP 150
Страница 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Страница 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Страница 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Страница 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Страница 27: ...Home screen Home screen The Home screen includes the following areas 23...
Страница 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Страница 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Страница 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Страница 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Страница 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Страница 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Страница 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Страница 133: ......