Electrical safety testing
Welch Allyn recommends performing ground continuity and leakage current tests after
all open-case repairs and dielectric strength testing
1
when replacing the power supply or
primary wiring according to EN/IEC 62353 - Medical Electrical Equipment - Recurrent
Test and Test After Repair of Medical Electrical Equipment.
Due to the variability of test equipment in the field, Welch Allyn does not include specific
instructions to perform electrical safety tests. When performing electrical safety tests,
refer to your test equipment manuals for detailed instruction. The following table
provides connections and test limits to assist you in performing these tests.
Test
Limits
Ground continuity
Ground continuity from EP stud* (equipotential terminal) to the Gnd pin of the
IEC power connector shall be no greater than 0.1 ohms.
Leakage current
Leakage current shall be less than 500 µA from EP stud* to mains (Line and
Neutral pins of the IEC power connector).
Insulation resistance
Insulation resistance measurements shall be performed with 500V DC from EP
stud to mains (Line and Neutral pins of the IEC power connector).
* To locate the equipotential terminal, see “Controls, indicators, and connectors.”
1
Perform this test only if there is a reason to doubt the integrity of the electrical insulation (e.g. multiple
trips of a residual-current device or liquid ingress of a saline solution). If you determine this test should
be performed, return the device to Welch Allyn for service.
58 Service
CP 150 12-lead resting electrocardiograph
Содержание CP 150
Страница 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Страница 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Страница 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Страница 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Страница 27: ...Home screen Home screen The Home screen includes the following areas 23...
Страница 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Страница 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Страница 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Страница 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Страница 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Страница 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Страница 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Страница 133: ......