1. Access the Advanced Settings.
a. Touch the
Settings
tab.
b. Touch the
Advanced
tab.
c. Enter
6345
as the
Advanced settings code
.
d. Touch
OK
.
The General tab appears at the bottom of the screen and the Regional tab appears at
the top of the screen.
2. Touch the
Network
tab.
The Status tab appears at the top of the screen.
3. Touch the vertical
Radio
or
Ethernet
tab to view the wireless Radio or Ethernet IP,
MAC address, and Status information.
4. Touch
(Next) to view more Ethernet or Radio information settings.
Information in the Status tab updates only when the device is connected to a wired
or wireless network.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home screen, touch
Exit
.
Specify Ethernet settings
You can connect a CP150 electrocardiograph to a Welch Allyn CardioPerfect workstation
or a network server via an Ethernet cable. Software provided in the CP 50/150
connectivity kit is required to communicate with the workstation. Cables longer than 3
meters have not been verified for use with the electrocardiograph. Do not use cables
longer than 3 meters.
1. Access the Advanced Settings.
a. Touch the
Settings
tab.
b. Touch the
Advanced
tab.
c. Enter
6345
as the
Advanced settings code
.
d. Touch
OK
.
The General tab appears at the bottom of the screen and the Regional tab appears at
the top of the screen.
If connecting the CP150 electrocardiograph to a Welch Allyn CardioPerfect
workstation, connect the electrocardiograph to the same network as the
workstation. If you need help, consult your network administrator.
2. Touch the
Network
tab.
3. Touch the
Ethernet
tab.
Service manual
Settings 31
Содержание CP 150
Страница 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Страница 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Страница 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Страница 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Страница 27: ...Home screen Home screen The Home screen includes the following areas 23...
Страница 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Страница 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Страница 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Страница 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Страница 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Страница 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Страница 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Страница 133: ......