Printer
Printer problems:
Symptom
Possible cause
Suggested action
The paper does not feed
The paper is jammed
Reload the paper.
Lost communication
Reseat the printer harness.
The printer door is open
Close the printer door.
The door sensor is faulty
Replace the door sensor.
The motor is faulty
Replace the motor.
The motor gear is faulty
Replace the gear assembly.
The paper is blank
Lost communication
Reseat or replace the printer harness.
Faulty print head
Replace the print head assembly.
Error loading the paper
Check to ensure that the printer door is
fully closed.
Check the orientation of the paper.
The wrong type of paper is being used Use only Welch Allyn approved paper.
The electrocardiograph shuts down
during printing when in battery mode
The battery is low or faulty
Recharge or replace the battery.
Printing is not clear or is garbled
Foreign particles on the printer heating
element
Clean the heating element of the
printer to remove any foreign particles.
Apply isopropyl alcohol to a cotton
swab and wipe off the heating element
surface. Allow to dry completely.
Faulty print head
Replace the print head assembly.
Loose platent roller
Reseat the platent roller on the paper
tray assembly.
Check the platent roller catch for a
crack. If a crack is found, replace the
paper tray assembly.
Door open message even when door is
closed
Loose door sensor harness connection Reseat door sensor harness connector.
Faulty door sensor
Replace door sensor.
Missing dowel pin on one side of the
paper tray assembly
Replace paper tray assembly.
Grinding sound in the gear
Wear of the gear assembly
Replace gear assembly.
64 Troubleshooting
CP 150 12-lead resting electrocardiograph
Содержание CP 150
Страница 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Страница 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Страница 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Страница 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Страница 27: ...Home screen Home screen The Home screen includes the following areas 23...
Страница 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Страница 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Страница 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Страница 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Страница 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Страница 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Страница 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Страница 133: ......