Remove and disassemble the printer assembly
•
Separate the top housing assembly from the bottom housing assembly as described.
1. Remove the ground cable at the J603 connection from the printer board (PCA).
2. Remove the four M4 pan head screws from the printer assembly using a # 2 Phillips
bit.
Reassembly notes
•
Use a #2 Phillips bit and a Torque driver calibrated to 5.0 in-lb ± 0.5 in-lb to secure
the printer assembly M4 pan head screws.
•
Route the ground cable around the outside of the screw boss during reassembly.
Service manual
Disassembly, repair, and reassembly 93
Содержание CP 150
Страница 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Страница 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Страница 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Страница 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Страница 27: ...Home screen Home screen The Home screen includes the following areas 23...
Страница 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Страница 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Страница 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Страница 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Страница 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Страница 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Страница 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Страница 133: ......