9. When the certificate update is complete, click
Close
.
10. For the Network Policy Certificate configuration to take effect, press and hold the
power button for about 8 seconds to reboot the device.
(Optional) Upgrade your CP 150 software using a USB
flash drive
Before you begin updating your CP 150, verify your current software version:
1. Touch the
Settings
tab.
2. Touch the
Advanced
tab. The software version appears on the Advanced Tab
screen. Take note of this version to verify that the latest software version is installed
when you reach the end of the upgrade process.
To perform the update, you will need the following:
•
A USB flash drive with the downloaded latest software.
•
A cable to connect the CP 150 to AC power.
•
An AC power outlet.
1. From the
Home
screen, touch the
Settings
tab.
2. Touch the
Advanced
tab.
3. Enter
6345
as the access code and touch
OK
.The General tab appears at the bottom
of the screen and the Regional tab appears at the top of the screen.
4. Touch the
Service
tab.
5. Touch
Update
to update the software and then touch
OK
to confirm the installation.
48 Service
CP 150 12-lead resting electrocardiograph
Содержание CP 150
Страница 1: ...CP 150 12 lead resting electrocardiograph Service manual Software version 2 10 XX...
Страница 8: ...4 Symbols CP 150 12 lead resting electrocardiograph...
Страница 22: ...18 Overview CP 150 12 lead resting electrocardiograph...
Страница 26: ...22 Controls indicators and connectors CP 150 12 lead resting electrocardiograph...
Страница 27: ...Home screen Home screen The Home screen includes the following areas 23...
Страница 30: ...26 Home screen CP 150 12 lead resting electrocardiograph...
Страница 42: ...38 Settings CP 150 12 lead resting electrocardiograph...
Страница 74: ...70 Guidance and manufacturer s declarations CP 150 12 lead resting electrocardiograph...
Страница 81: ...Cable identification for reassembly Service manual Disassembly repair and reassembly 77...
Страница 110: ...106 Disassembly repair and reassembly CP 150 12 lead resting electrocardiograph...
Страница 128: ...124 Partners in Care service and support agreements CP 150 12 lead resting electrocardiograph...
Страница 132: ...Interconnect diagram 128 Appendices CP 150 12 lead resting electrocardiograph...
Страница 133: ......