Final check
13
3.2 Preparation for the final check
1. Device setting freq. = 30 min
-1
, V
t
= 0.1 l
2. Check which language is set. Record setting.
3. Switch on audio response (see “2.5 Audio response” on Page 9).
4. Check whether the metronome is on or off (see “2.6 Switching the metronome on and off” on Page 11).
Record setting.
3.3 Enter device data
1. Enter the device data from the device information plate in the test protocol and check the device
information plate is complete and undamaged.
3.4 Check for leaks and check pressure display
3.4.1 Check for leaks on inlet side
1. Provide a pressurized gas supply to MEDUMAT Easy CPR and to the flow meter.
2. Switch off device.
3. Open oxygen cylinder.
4. Close oxygen cylinder.
Requirement:
Drop in pressure must be less than 0.2 bar/min.
13
14
0,3
0,6
1,1
60
50
40
30
20
10
0
CPR