Repair information and instructions
29
6. Repair information and instructions
6.1 General
Always perform repairs to MEDUMAT Easy CPR at an ESD-protected workplace.
•
Observe the safety information in the Operating
Instructions for MEDUMAT Easy CPR.
• All operations on this device require detailed
knowledge and observation of the Operating
Instructions and the Service and Repair
Instructions.
• Do not carry out any repairs that are not
described in these Service and Repair
Instructions. This is the only way to guarantee
trouble-free functioning of
MEDUMAT Easy CPR.
• Make sure that your hands and workplace are
clean during the repair work.
• Be sure to carry out a final check after every
repair (see "3. Final check" on Page 8).
• If you replace components or individual parts,
use only genuine WEINMANN parts.
• When ordering the lower part of the
, please specify device type, year
of manufacture and device number.
•
Note:
The item numbers quoted in the following text
are identical to the item numbers in the spare
parts list on page 35 and the overview on
page 4.
6.2 Changing the filter in the pressurised gas connection
Tools and equipment required
• Flat-head screwdriver
• Tweezers.
1. Unscrew the slot-head screw
2. Use tweezers to remove filter set
3. Carefully insert a new filter set
pressurised gas connection.
4. Screw the slot-head screw
firmly into the
pressurised gas connection
.