32
Repair information and instructions
6.4 Close the device
Materials and tools required
• ESD-protected workplace
• Torque wrench 50 ± 5 Ncm,
• Torque wrench 200 ± 10 Ncm,
• Tubular hexagon box wrench, 10 mm
• Special tool WM 22829 from special tool set WM 15349
• Hex socket wrench, size 2.5
1. Place the two halves of the housing next to one
another with the outsides face down.
2. If the Y sensor tube was damaged: push tube
Y sensor tube (tube set WM 15058) together
with Y connector
3. Put the battery contacts in the battery
compartment and screw them with the two
screws
Note:
The shortest tube 59 of the Y sensor tube may
only be pushed onto sensor S3003, otherwise it
may kink!
4. Push tube
(the shortest of the three tubes of
the Y sensor tube) carefully onto sensor
S3003
S3000
on the board.
5. Screw the long tube
Y sensor tube to the pneumatic block.
6. Screw the ground connection to the pneumatic
and serrated lock
washer
.
7. Connect long tube
.
8. Screw the sensor tube to the pneumatic block.
9. Connect the connector for cable harness
terminal
X1001
on the board.
10. Connect the connector for the MEDUtrigger to
terminal
X1003
on the board.
S3003
S3000
X1003
X1001
Ground connection