Test record
as per Servicing and Repair Instructions WM 67308 (Medumat Easy CPR)
Geräte für Medizin GmbH+Co. KG, P.O. Box 54 02 68, D-22502 Hamburg, Germany, Fax +49 4054 702461,
Tel49 4054 7020
WM 28205 - 0
4/
10
Page 3 of 4
3.9.
Function check of frequency setting
OK
NOK
•
Frequency 14 ± 2 min
-1
min
-1
•
Frequency 10 ± 2 min
-1
min
-1
•
Frequency 30 ± 2 min
-1
min
-1
3.10.
Function check of breath volume at 4.5 bar input pressure and 10 mbar back
pressure
OK
NOK
•
Breath volume 1140 ± 170 ml
ml
•
Breath volume 1100 ± 170 ml
ml
•
Breath volume 100 ± 20 ml
ml
3.11.
Test O
2
concentration
OK
NOK
•
O
2
concentration must be > 98 % by volume
3.12.
Function check of pressure limit
OK
NOK
•
Pressure limit 20 ± 5 mbar
mbar
•
Pressure limit 45 ± 5 mbar
mbar
3.13.
Function check of venting valve without patient valve
OK
NOK
•
The ventilator vents audibly
3.14. Test the MEDUtrigger
OK
NOK
•
LEDs on Medutrigger come on
•
LEDs on Medutrigger go out
3.15. Test communication socket
OK
NOK
•
The communication test runs without errors
(green 30 mbar LED comes on)
•
The communication test runs without errors
(red 60 mbar LED comes on)
3.16. Check external condition
OK
NOK
•
Housing not scratched and has no defects
•
Connecting thread undamaged and working smoothly
•
Knob inhibited to prevent unintentional adjustment
•
Elbow outlet is easy to rotate.