Repair information and instructions
31
8. Take the connector for the MEDUtrigger off
terminal
X1003
on the board.
9. Detach the connector for cable harness
from terminal
on the board.
Note:
This is most easily achieved using flat-nose
pliers. However, under no circumstances
damage the board with the flat-nose pliers.
10. Unscrew the sensor tube from the pneumatic
block.
Note:
The sensor tube may not be pulled by the
pressure sensor on the board and this damages
it!
11. Separate the long tube
Y sensor tube from the Y connector
with the
12. Pull the two halves of the housing apart and put
them down with the outsides face down.
13. Undo screw
from the ground connection to the pneumatic
block.
14. If the Y sensor tube is damaged: carefully take
off sensor
S3003
and tube
off
sensor
S3000
on the board.
15. Unscrew the long tube
Y sensor tube from the pneumatic block.
16. Undo the two screws
and pull the battery contacts out of the battery
compartment.
Both halves of the housing are now separated
from one another.
X1003
X1001
Sensor tube
S3003
S3000
Ground connection
S3003
S3000