Test record
as per Servicing and Repair Instructions WM 67308 (Medumat Easy CPR)
Geräte für Medizin GmbH+Co. KG, P.O. Box 54 02 68, D-22502 Hamburg, Germany, Fax +49 4054 702461,
Tel49 4054 7020
WM 28205 - 0
4/
10
Page 2 of 4
Preparation
3.1.
Testing equipment
(
as per Servicing and Repair Instructions WM 67308)
3.2.
Preparation for the final check
(
as per Servicing and Repair Instructions WM 67308)
3.3. Enter device data
Tests
3.4. Check for leaks and check pressure display
OK
NOK
3.4.1. Check for leaks on inlet side
•
Drop in pressure is less than 0.2 bar/min
3.4.2. Check for leaks in pressure measuring section
•
Drop in pressure is less than 2 bar/min
3.5.
Self-test after device is switched on
OK
NOK
•
The self-test runs without errors
3.6.
Function check of controls (key test)
OK
NOK
•
All the controls work perfectly
3.7.
Function check and alarms
OK
NOK
3.7.1. Test stenosis alarm
•
The stenosis alarm is correctly triggered
3.7.2. Test alarm acknowledgement
•
The acoustic alarm is suppressed immediately, the visual alarm remains active
3.7.3. Test disconnection alarm
•
The disconnection alarm is correctly triggered
3.7.4. Test pressure alarm
•
The pressure alarm is correctly triggered
3.7.5. Test demand mode LED
•
The LED must come on/flicker
3.7.6. Test audio response
•
The audio response must be clear and intelligible.
3.8.
Battery voltage
OK
NOK
3.8.1 3.0 V battery (board)
•
Battery voltage is between 3.0 and 3.2 V
V
3.8.1 3.6 V battery (battery compartment)
•
Battery voltage is between 3.4 and 3.7 V
V