3.4.2 Check for leaks in pressure measuring section
1. Put T-tube connection with syringe (WM 15359) on the pressure measurement tube connection
2. Use the syringe to generate a pressure of 55 mbar ± 2 mbar on the test pressure gage.
Requirement:
Drop in pressure must be
≤
2 mbar/min.
3.5 Self-test after the device is switched on
1. Provide a pressurized gas supply to MEDUMAT Easy CPR and to the flow meter.
2. Connect the patient circuit to MEDUMAT Easy CPR.
3. Open oxygen cylinder.
4. Switch on device.
Requirement:
The self-test is triggered, the ventilation mode LEDs light up once each in turn, the alarm
LEDs flash, the pressure indicator LEDs are run through 3x, a signal sounds and one sentence is issued
by the audio response.