June 11, 2015, 715004752 Rev. A
Page 118
Figure 5–8:
Contact closures & events diagnostic display
The Input & Output diagnostic test allows real-time monitoring of the state of
the contact closure inputs. A solid (filled in) circle indicates the contact closure
is closed (ON = High). An open (empty) circle indicates the contact closure is
open (OFF = Low).
2. For the outputs (SW1 and SW2):
a. Press Enter to display the active switch (surrounded by a dotted-line
border).
b. Press any numerical key to change the status of the switch (ON to OFF or
vice versa).
c.
Press Enter to select the second switch.
5.2.2.6
Using the lamp, display, and keyboard diagnostic tests
To access the lamp, display, and keypad diagnostic tests, press DIAG and then press
4.
Requirement:
Make sure the detector is powered-off and the power cord is
disconnected before you replace the lamp.
To change the lamp:
1. From the Lamp, display & keypad choice list, press 1 (Change lamp) to access
the Change Lamp diagnostic display.
Figure 5–9:
Change Lamp diagnostic display
2. Enter the 9-digit serial number and date of installation for the new lamp. Press
Enter after each entry.
Tip:
Use this function whenever you change the lamp to enter a new serial
number and installation date (see
explanation of the lamp replacement procedure).
Important:
Neglecting to record the new lamp serial number following the
prescribed procedures referenced above voids the lamp warranty. The date of
the previous lamp installation remains in the detector memory.
Содержание 2489
Страница 2: ...June 11 2015 715004752 Rev A Page ii...
Страница 8: ...June 11 2015 715004752 Rev A Page viii...
Страница 14: ...June 11 2015 715004752 Rev A Page xiv...
Страница 46: ...June 11 2015 715004752 Rev A Page 46...
Страница 124: ...June 11 2015 715004752 Rev A Page 124...
Страница 135: ...June 11 2015 715004752 Rev A Page 135 Lower limit of temperature Temperature limitation Symbol Description...
Страница 136: ...June 11 2015 715004752 Rev A Page 136...