June 11, 2015, 715004752 Rev. A
Page 107
Figure 4–7:
Removing the lamp assembly
4.5.4 Installing the new lamp
Important:
Do not touch the glass bulb of the new lamp. Dirt or fingerprints on the
bulb adversely affect detector operation. If the lamp needs cleaning, gently clean the
bulb with ethanol and lens tissue. Do not use abrasive tissue or apply excessive
pressure.
Before you begin:
1. Unpack the lamp from its packing material.
Note:
The new lamp’s features can vary slightly from those shown in the
figure.
2. If you see any particles or dirt on the new lamp, clean the lamp with a gas
duster or lens tissue.
3. Record the serial number, located on a label attached to the lamp connector
wire, using the procedure in
“Recording the new lamp serial number” on
Requirement:
Make sure the detector is powered-off and the power cord is
disconnected.
Note:
When you change the lamp, always power-off the detector. After installing a
new lamp, power-on the detector and always wait at least 5 minutes for the new lamp
to warm up.
Warning:
To avoid exposing your eyes to harmful ultraviolet radiation, never
ignite a lamp while it is outside the instrument or if it is not properly secured in
place.
Lamp assembly
Содержание 2489
Страница 2: ...June 11 2015 715004752 Rev A Page ii...
Страница 8: ...June 11 2015 715004752 Rev A Page viii...
Страница 14: ...June 11 2015 715004752 Rev A Page xiv...
Страница 46: ...June 11 2015 715004752 Rev A Page 46...
Страница 124: ...June 11 2015 715004752 Rev A Page 124...
Страница 135: ...June 11 2015 715004752 Rev A Page 135 Lower limit of temperature Temperature limitation Symbol Description...
Страница 136: ...June 11 2015 715004752 Rev A Page 136...