June 11, 2015, 715004752 Rev. A
Page 91
Recommendation:
Program the lamp to shut off or turn the lamp off manually
only if the value of the “Lamp off” parameter is more than 4 hours.
Without powering off the system, you can conserve lamp life by,
•
turning power to the lamp off and on again manually;
•
programming a timed event to turn the lamp off and on again;
•
programming the lamp to turn off and on again using the external contact
closure.
To turn power to the lamp on and off manually, use the Lamp keypad function (Shift
1). When the lamp is turned off, the absorbance screen displays the words “Lamp off”,
and the lamp icon appears with an “X” through it.
Use the Lamp key (Shift 1) to,
•
shut the lamp off or turn the lamp on manually;
•
display the number of times the lamp has ignited;
•
display the hours and minutes the lamp has been turned on during the current
run, since its installation, or both.
To turn off the lamp manually from the detector front panel:
1. Press Lamp (Shift 1) from the keypad.
Result:
The lamp control screen appears.
Figure 3–34: Lamp control screen
2. Press Lamp (Shift 1) again to turn the lamp off.
Figure 3–35: Lamp off/on sequence
To ignite the lamp manually:
1. When there is an X through the lamp icon on the absorbance screen, press
Lamp (Shift 1).
2. Press Lamp (Shift 1) again to turn the lamp on.
Lamp off indicator
Lamp on indicator
Содержание 2489
Страница 2: ...June 11 2015 715004752 Rev A Page ii...
Страница 8: ...June 11 2015 715004752 Rev A Page viii...
Страница 14: ...June 11 2015 715004752 Rev A Page xiv...
Страница 46: ...June 11 2015 715004752 Rev A Page 46...
Страница 124: ...June 11 2015 715004752 Rev A Page 124...
Страница 135: ...June 11 2015 715004752 Rev A Page 135 Lower limit of temperature Temperature limitation Symbol Description...
Страница 136: ...June 11 2015 715004752 Rev A Page 136...