Sterilization cycles
Load type
Placement
Pouched items
On trays allowing adequate space in-between
bags. Ensure that packs do not touch the walls of
the chamber. Place sterilization pouched items
with the paper side facing up.
Items made from different
materials (stainless steel,
carbon steel, aluminium, etc.)
On separate trays or wrapped/pouched
Instruments manufactured
from carbon steel
Place paper among them and the trays to avoid
rusty spots
PARTIAL LOAD
If the chamber is just partially loaded, place the load in such a way
that the space in-between the trays is maximized. Spread items
evenly on multiple trays. Below is a example with three trays.
Typical loads
EXAMPLES
An example of a typical load used during the tests performed by
W&H is reported below: