Lexa PLUS | Instructions for Use | RIS-311 ENG Rev03 | 29/08/2022 | © W&H Sterilization Srl
61
Sterilization cycles
TEST THE DATA STORAGE
Note
: The test function is available only if the remote data storage is
enabled. See "Set the remote folder for saving the reports" on the previous
page.
1
On the homepage, tap
>
>
2
Tap : a sequence of tests is automatically performed.
3
In case a test fails, check the relevant settings and tap
to
repeat the test sequence, if the error persists call your IT
manager.
4
Tap to go back to the previous page.
SAVE ALL THE CYCLE REPORTS IN A REMOTE FOLDER
Note
: The save all function is available only if the remote data storage is
enabled. See "Set the remote folder for saving the reports" on the previous
page.
Only the last 400 cycles in HTML and 5 cycles in SCL in the sterilizer
memory can be saved in the remote folder.
1
On the homepage, tap
>
>
>
2
Tap
to start the remote saving.