Sterilizer tests
46
Lexa PLUS | Instructions for Use | RIS-311 ENG Rev03 | 29/08/2022 | © W&H Sterilization Srl
DESCRIPTION
It consists of several sheets of paper wrapped in a small packet with a chemical heat-sensitive indicator sheet in the middle. The colour assumed
by this indicator sheet at the end of the sterilization cycle gives the result of the test.
CARRY OUT THE TEST
Note
: carry out the test according to the local Regulations and the instructions of the PCD manufacturer.
1
Empty the sterilizer chamber to ensure no load is present. Remove all trays from
the sterilization chamber, except the lowest one.
2
Place the Bowie and Dick test pack in the center of a tray in the lowest rack
position and close the chamber door.
3
On the homepage, tap
B&D test
.
4
To set the duration of the Plateau/Sterilization phase and other settings, tap
.
5
Tap and enter your credentials if required: the chamber door locks.
6
Wait until the end of the test and tap
OPEN
: the chamber door unlocks.
7
Enter your credentials if required.