Sterilizer tests
Lexa PLUS | Instructions for Use | RIS-311 ENG Rev03 | 29/08/2022 | © W&H Sterilization Srl
47
8
Open the chamber door, extract the tray using the tray holder and take the test
pack. The test pack can be wet outside.
CAUTION!
Risk of burns. The test pack is very hot at the end of the cycle. Wear
appropriate PPE (e.g. gloves).
9
Remove the indicator sheet from the center of the test pack and check the change
in colour. See "Bowie and Dick test" on page 45
Note
: for interpreting the test result follow the instructions of the PCD manufacturer.
Vacuum test
CAUTION!
Follow local/national guidelines on the frequency of testing.
Notice
: If a drainage period of the Unwrapped cycle is still operating, wait the drainage to be finished and the sterilizer to be both cold and dry.
Otherwise, a false negative outcome could occur.
PURPOSE OF THE TEST
The test is used to validate the sterilizer performance in terms of:
n
efficiency of the vacuum pump
n
tightness of the pneumatic circuit
DESCRIPTION
It consists of a vacuum phase, followed by a stabilization period of 5 minutes and a testing period of 10 minutes. The internal pressure is
monitored during the testing period. The pressure rise must be less than 0.013 bar (0.19 psi).