Technical data
92
Lexa PLUS | Instructions for Use | RIS-311 ENG Rev03 | 29/08/2022 | © W&H Sterilization Srl
Sterilization cycles
Wrapped
Low Temperature
Unwrapped
1
Sterilization temperature
270 °F (132 °C)
250 °F (121 °C)
270 °F (132 °C)
Sterilization program type
according ANSI/AAMI ST55
Dynamic-air-removal:
Pre-vacuum
Gravity displacement
Dynamic-air-removal:
Pre-vacuum
Sterilization program type
according EN 13060
Class B
Class B
Class S
Sterilization time (minutes)
4'
30'
4'
Drying time (minutes)
Range: 5' – 22'
Recommended: 22'
Range: 5' – 28'
Recommended: 28'
Range: 0' – 4'
Recommended: 4'
Total program duration (min)
2
Full loaded including drying time
Empty: 28'
Full: 62'
Typical: 39'
Empty: 68'
Full: 107'
Typical: 85'
Empty: 20'
Typical: 24'
Load type
All types of load
*
All types of load except handpieces
*
All types of load except porous/textile
*
Maximum load weight
Pouched/unwrapped items on trays: 14 lbs
(6.3 kg)
3
Items in wrapped/unwrapped cassettes or in
trays and wrapped/unwrapped cassettes: 16.5
lbs (7.5 kg)
4
Pouched/unwrapped textile load on trays: 4.4
lbs (2.0 kg)
3
Pouched/unwrapped items on trays: 14 lbs
(6.3 kg)
3
Items in wrapped/unwrapped cassettes or in
trays and wrapped/unwrapped cassettes: 16.5
lbs (7.5 kg)
4
Pouched/unwrapped textile load on trays: 4.4
lbs (2.0 kg)
3
Unwrapped items: 18 lb (8.1 kg)
3
*
: the liquids are excluded.
1
: the instruments sterilized with this cycle cannot be stored: they must be used immediately after being sterilized.
2
: the total cycle time may vary depending on the type of load (solid or porous), the load weight, the duration of the drying phase (if the ECO DRY plus mode is
disabled) and other factors. Values and cycle names could be different depending on country requirement.
3
: the wheight of trays is excluded.
4
: the wheight of trays and cassettes is included.