12
5. Safety notes
Please observe the following instructions under all circumstances
> Before using the Implantmed for the first time, store it at room temperature for 24 hours.
> Only fit the straight and contra-angle handpieces when the motor is at a complete standstill.
> Never touch rotary instruments that are still rotating.
> Never touch the chuck mechanism of straight and contra-angle handpieces during operation or while they are still
running down.
> Always ensure correct operating conditions and that sufficient and adequate cooling is delivered.
> Avoid overheating at the treatment site.
> Check the Implantmed, the straight or contra-angle handpiece and the motor with cable for damage and loose parts
each time before using. Correct any faults or refer to an authorized W&H service partner (see page 52). Do not operate
the Implantmed if it is damaged.
> When changing the fuse, disconnect the unit from the power supply and only use W&H original fuses.
> Perform a test run prior to each treatment.
> Never touch the patient and the connections for the foot control simultaneously.
> Check the parameter settings every time the device is restarted.
> The ESD spring contact on the underside of the foot control must touch the floor during operation.
Use only suitable and serviceable tools
Ensure that you comply with the manufacturer’s instructions for surgical straight and contra-angle
handpieces with respect to maximum speed, maximum torque, forward and reverse movement.
Inappropriate use
Improper use, in addition to incorrect assembly, installation, modification or repairs of the Implantmed or
failure to comply with our instructions invalidates all claims under warranty and any other claims.