44
20. Technical data
Implantmed
SI-923 SI-915
Supply voltage:
220 –240 V
100 –130 V
Permitted voltage fluctuation:
± 10 %
± 10 %
Nominal current:
0.1 –0.8 A
0.2 –1.7 A
Frequency:
50 –60 Hz
50 –60 Hz
Mains fuse:
2x 250 V – T1.25AH
2x 250 V – T1.25AH
Max. power consumption:
170 VA
170 VA
Max. mechanical output power:
70 W
70 W
Max. torque on the motor:
5.5 Ncm
5.5 Ncm
Motor speed range in the nominal voltage range:
300 – 40,000 min.
-1
300 – 40,000 min.
-1
Coolant flow rate at 100%:
at least 90 ml/min
at least 90 ml/min
Operating mode:
S3 (3min / 10min)
S3 (3min / 10min)
Dimensions in mm (WxDxH):
235 x 240 x 100
235 x 240 x 100
weight control-unit with motor in kg:
2.7
2.7
Physical characteristics
Temperature for storage and transport: -40 °C to +70 °C
Humidity for storage:
8 % to 80 % (relative), non-condensing at +40 °C
Temperature during operation:
+10 °C to +40 °C
Humidity during operation:
15 % to 80 % (relative), non-condensing at +40 °C