32
15. Thread cutter function (chip breaker mode)
Press program button P5.
Use Plus/Minus to increase or decrease the torque.
When the thread cutter function (P5) is activated, the speed in both forward and reverse operation modes is 20 rpm and
can no longer be changed. When the foot pedal (grey) on the foot control is pressed, the thread cutter rotates inwards
until the set torque is reached. When the set torque is reached, the device automatically switches to reverse operation.
Disengaging and then re-engaging the foot pedal will switch the device back to forward operation.
If the thread cutter function is in reverse operation mode, the device can also start with the maximum torque.