Safety notes
Coolant
The Implantmed is designed for use with physiological saline solution. Use only suitable irrigation fluids and follow the
manufacturer’s medical data and instructions. Use the W&H irrigation tubing set or accessories approved by W&H.
You can purchase the coolant bottle or the coolant bag at a drugstore.
Sterility of irrigation tubing set
Sterile irrigation tubing sets are supplied with the equipment. The irrigation tubing is a disposable article and must be
discarded after each treatment. Please note the expiry date and the relevant regulations for disposal of irrigation tubing.
Only use disposable irrigation tubing with undamaged packaging.
Rotational energy
As a result of the rotational energy stored in the drive system – compared to the value set – fast deceleration of the bur can
cause the selected torque to be significantly overloaded, at times.
Observe the individual manufacturer’s instructions for use when adjusting superstructure screws. We would
point out that adjusting these screws with an electric motor presents a potential risk as described above.