16. Error messages
Error no. Description
Solution
00
Electronics overheating–Safety shutdown
Switch off device, allow device to cool for at least
10 minutes, re-start
01
Electronics overloaded
Switch off device, allow device to cool for at least
10 minutes, re-start
07
Foot control error –
initializing
Switch off device, re-start, do not actuate
foot control when switching on
09
Foot control error
Switch off device, check plug-in
connection of foot control, re-start
19
Running time limiter
Switch off device and re-start
99
System failure
Switch off device, allow motor to cool for at least
10 minutes, re-start
Motor temperature too high
Motor plug-in connection
Switch off device, check plug-in motor connection,
allow motor to cool for at least 10 minutes, re-start
If one of the error messages described above cannot be rectified by switching off the Implantmed and then switching
it on again, the equipment must be checked by an authorized W&H service partner (see Page 52). If a total failure of
the equipment occurs caused by external circumstances, the equipment must be switched off and then on again.