25
Control unit operation – change torque (P4 – P5)
Program P4:
range from 5 – 70 Ncm, intermediate stage 32 Ncm.
Program P5:
range from 20 – 60 Ncm.
Pressing and holding PLUS / MINUS depressed activates the repeat function
and the values are continuously increased / decreased.
A longer confirmation signal is heard on changing from 5 to 70 Ncm (20 to 60) and 70 to 5 Ncm (60 to 20).
The motor switches off automatically when the set torque is reached in forward and reverse operation modes.
Press program button P4 or P5
P4:
Increase torque in 5-Ncm steps
P5:
Increase torque in 10-Ncm steps
P4:
Decrease torque in 5-Ncm steps
P5:
Decrease torque in 10-Ncm steps
The accuracy of the set torque at 50 Ncm is ± 10% with the WI-75 and WS-75 contra-angle handpieces.