37
Hygiene and maintenance
Motor with cable
Mechanical cleaning and disinfection internal and external
The motor with cable can be cleaned and disinfected in a thermo washer disinfector.
W&H permits preparation in a thermo washer disinfector with a drying program.
> Follow the manufacturer’s recommendations for devices, cleaning and rinsing agents.
Ensure that the motor with cable is completely dry internally and externally after thermo washer disinfection.