15
Safety notes
Mains cable
Only use the mains cable supplied.
Only connect to a grounded socket outlet.
Set up the device so that the power switch is easily accessible.
In dangerous situations, the device can be disconnected from the power supply using the power switch or power cable.
The power switch can also be used to safely stop the device.
Power failure
In the event of a power failure, if the Implantmed is switched off, or when switching between programs, the last values set are saved
and re-activated on power-up.
System failure
A total system failure does not constitute a critical fault.
Simply switch the unit off and then on again.
Intermittent operating mode S3 (3 min/10 min)
The Implantmed is designed for intermittent operating mode S3 with an operating time of 3 minutes and an idle time of 10
minutes.If the operating mode specified is observed no overheating of the system and therefore no injury to the patient,
user or third persons arises. The responsibility for the use and timely shutdown of the system lies with the user.