VTI Instruments Corporation
94
VM2710A Command Dictionary
CONFigure?
Purpose
Reports the most recent configuration set
Type
Query
Command Syntax
N/A
Command Parameters
N/A
*RST Value
N/A
Query Syntax
CONFigure?
Query Parameters
N/A
Query Response
Current function, range, and resolution parameters
Description
The CONfigure? query reports the multimeter's current configuration set be the most
recent Configure or Measure command and returns these values in quotation marks.
Examples
Command / Query
Response (Description)
CONF:VOLT:AC 100,0.001
CONF?
“VOLT:AC 1.002,1.000000E-003”
(Indicates that the device is con figured for volts
ac in the 100 V range at a resolution of 1 mV.)
Related Commands
CONFigure:CURRent:AC
CONFigure:CURRent[:DC]
CONFigure:VOLTage:AC
CONFigure:VOLTage[:DC]
CONFigure:FRESistance
CONFigure:RESistance
MEASure:CURRent:AC
MEASure:CURRent[:DC]
MEASure:VOLTage:AC
MEASure:VOLTage[:DC]
MEASure:FRESistance
MEASure:RESistance