www.vtiinstruments.com
VM2710A Command Dictionary
115
MFGTEST 99
Purpose
Enables or disables the ohms current function
Type
Setting
Command Syntax
MFGTEST 99 <boolean>
Command Parameters
<boolean> = ON | OFF | 1 | 0
*RST Value
1
Query Syntax
MFGTEST? 99
Query Parameters
N/A
Query Response
0 | 1
Description
The MFGTEST 99 command enables or disables the ohms current function. This can
be used to help measure resistance in a circuit with other currents flowing. The default
setting is enabled.
Examples
Command / Query
Response (Description)
MFGTEST 99 0
(Disables the ohmes current function.)
MFGTEST? 99
0 (Verifies that the ohms current function is disabled)
Related Commands
N/A