VTI Instruments Corporation
134
VM2710A Command Dictionary
[SENSe:]FUNCtion
Purpose
Selects the measurement function
Type
Setting
Command Syntax
[SENSe:]FUNCtion:<function type>
Command Parameters
<function type> = FRESistance | RESistance | VOLTage:AC | VOLTage[:DC]
*RST Value
VOLTage[:DC]
Query Syntax
[SENSe:]FUNCtion?
Query Parameters
N/A
Query Response
FRES | RES | VOLT:AC | VOLT
Description
This command selects the measurement function of the multimeter. The default
function is dc voltage.
Examples
Command / Query
Response (Description)
FUNC:RES
(Selects the 2-wire ohms function.)
FUNC?
RES
Related Commands
N/A