VM2710A Command Dictionary
MEASure:RESistance?
Purpose
Selects the 2-wire ohms function
Type
Query
Command Syntax
N/A
Command Parameters
N/A
*RST Value
DEFault - Autoranging / 1 PLC
Query Syntax
MEASure:RESistance? [<expected value>[, <resolution>]]
Query Parameters
<expected value>
= 0 Ω through 20 MΩ | AUTO | DEF | MIN | MAX
<resolution>
= 1e-4 | 1e-3 | 1e-2 | 1e-1 | 1 | 10 | 100 | 1000 | DEF | MIN | MAX
Query Response
numeric
Description
This command selects the 2-wire ohms function. It also allows the user to specify an
expected value and desired resolution. When a specific resistance value is specified,
the multimeter will automatically selects the correct range. If autoranging is selected,
the only valid parameters for resolution are
MINimum
or
MAXimum
. To input a
resolution parameter, a resistance parameter must be specified first. Valid parameter
combinations are: no parameter, a resistance parameter only, or both a resistance
parameter and a resolution parameter. See the
Resistance Specifications
in Section 1
for allowed range and resolution values. The input parameters are as follows:
Resistance
0 Ω - 20 MΩ
Specific expected resistance value
AUTOmatic
Autoranging
DEFault
Autoranging
MINimum
0 Ω
MAXimum
20 MΩ
[no value specified]
Autoranging
Resolution
1e-4 | 1e-3 | 1e-2 | 1e-1 | 1 | 10 | 100 | 1000
A value specified for Resolution
DEFault
1 Ω
MINimum
(Dependant on Range)
MAXimum
(Dependant on Range)
[no specified value]
1 Ω (Dependant on Range)
Examples
Command / Query
Response (Description)
MEAS:RES MAX,MAX
(Selects 20 MΩ at 100 Ω resolution.)
MEAS:RES 10,0.001
(
Selects the 20 Ω range at 0.001 Ω resolution.)
Related Commands
READ?