VTI Instruments Corporation
80
VM2710A Command Dictionary
CALibration:CONVert:OFFSet:CURRent[:DC]?
Purpose
Reports the calibration offset value
Type
Query
Command Syntax
N/A
Command Parameters
N/A
*RST Value
N/A
Query Syntax
CALibration:CONVert:OFFSet:CURRent:DC? <range>
Query Parameters
<range> = 0.01 | 0.1 | 1.0
Query Response
See Table 3-1: Function and Range
Description
Reports the offset for a specified range in the dc current function. See the Offset
column in Table 3-1: Function and Range.
Calibration commands should only be executed by qualified
personnel. Changing these values incorrectly can cause the
instrument to perform improperly.
Examples
Command / Query
Response (Description)
CAL:CONV:OFFS:CURR:DC 1
(Queries and reports the offset for the 1 A
range.)
Related Commands
CALibration:CONVert:OFFSet?
CALibration:CONVert:OFFSet:CURRent:AC?
CALibration:CONVert:OFFSet:FRESistance?
CALibration:CONVert:OFFSet:RESistance?
CALibration:CONVert:OFFSet:VOLTage:AC?
CALibration:CONVert:OFFSet:VOLTage[:DC]?