VTI Instruments Corporation
36
VM2710A Programming
CALibration:LFRequency?
The query will report the current frequency setting of
50 Hz or 60 Hz. If MINimum or MAXimum is
specified, it will report the values available, which
are 50 Hz or 60 Hz, respectively.
CALibration:NUMBer?
The Calibration Number is the number of times the
device has been calibrated. This number is
incremented each time the CAL:STORE command is
issued to the device.
CALibration:RESet
This command will load the Calibration DACs from
the calibration constants stored in non-volatile
memory. This allows the user to modify Calibration
DAC values with the ability to restore the original
values. Note that CAL:SEC:STAT must first be OFF.
CALibration:SECure:CODE
This command/query sets or queries the
CALibration:SECure:CODE. Both the command and
query require the CALibration:SECure:STATe to be
OFF. There is no driver function for this
command/query.
CALibration:SECure[:STATe]
This command disables or enables the Calibration
Secure state. This command is NOT dependant on
any other settings on the VM2710A other than the
CALibration:SECure:CODE. This command is
performed by the driver function
vtvm2710_enableDisableSecState.
CALibration:STORe
This command stores the currently loaded calibration
values into Non-volatile memory. Before this
command is issued, the instrument will use the
currently loaded values, but will not save these
values after power is removed, or a reset is issued.
The vtvm2710_storeCalibrationData driver function
performs this command. Note that CAL:SEC:STAT
must first be OFF.
CALibration:ZERO
Sets the range for the zero measurement in the dc
voltage function and takes a zero reading. If no
value is entered, then the VM2710A will take a zero
measurement for each of the voltage ranges, thus
taking more time.
CALibration:ZERO:UPDate
Updates the current zero measurements to later be
stored using the Calibration Store command.
See the
Command Dictionary
section for more information on the Calibration commands.