VTI Instruments Corporation
16
VM2710A Introduction
AC
V
OLTAGE
S
PECIFICATIONS
M
EASUREMENT
M
ETHOD
AC-coupled true rms
AC
R
ANGES
0.1, 1.0, 10.0, 100.0, 300.0 VAC
O
VERRANGE
20% except 300 V range
I
NPUT
I
MPEDANCE
1 MΩ in parallel with < 50 pF
C
REST
F
ACTOR
5:1 maximum
A
DDITIONAL
C
REST
F
ACTOR
E
RROR
Crest Factor
1 - 2
2 - 3
3 - 4
4 - 5
Add 0.05% of reading
Add 0.15% of reading
Add 0.30% of reading
Add 0.40% of reading
AC
C
OMMON
-
MODE
R
EJECTION WITH
100
Ω
IN
LOW
L
EAD
70 dB
M
AXIMUM
I
NPUT
V
OLTAGE
dc + ac = 300 V
rms
R
ESOLUTION
Range
100 PLC
(6.5 Digits)
10 PLC
(6.5 Digits)
1 PLC
(5.5 Digits)
0.1 PLC
(5.5 Digits)
0.03 PLC
(4.5 Digits)
0.1 V
1
µ
V
1
µ
V
1
µ
V
10
µ
V
10
µ
V
1.0 V
10
µ
V
10
µ
V
10
µ
V
100
µ
V
100
µ
10.0 V
100
µ
V
100
µ
V
100
µ
V
1 mV
1 mV
100.0 V
1 mV
1 mV
1 mV
10 mV
10 mV
300.0 V
1 mV
1 mV
1 mV
10 mV
10 mV
A
CCURACY
@
100,
10,
AND
1
PLC
FOR
24
HR
@
T
CAL
±1
°C
AFTER
1
HR WARM
-
UP
;
±(%R
EADING
+
%R
ANGE
):
S
INE
-
WAVE INPUT
≥
5%
F
ULL
S
CALE
(F.S.),
L
OW
-F
REQUENCY
F
ILTER
ON
Range
0.1 V Range
1.0 V – 300.0 V Ranges : (V
IN
x Freq.
≤
1.5x10
7
)
20 Hz – 30 Hz
1.00 + 0.15
1.00 + 0.02
30 Hz – 50 Hz
0.50 + 0.15
0.50 + 0.02
50 Hz – 200 Hz
0.20 + 0.15
0.20 + 0.02
200 Hz – 10 kHz
0.05 + 0.15
0.02 + 0.02
10 kHz – 50 kHz
0.20 + 0.15
0.05 + 0.04
50 kHz – 100 kHz
0.60 + 0.20
0.30 + 0.10
100 kHz – 300 kHz
3.00 + 0.50
3.00 + 0.50
A
CCURACY
@
100,
10
AND
1
PLC
FOR
90
DAYS
@
T
CAL
±5
°C
AFTER
1
HR WARM
-
UP
;
±(%R
EADING
+
%R
ANGE
):
S
INE
-
WAVE INPUT
≥
5%
F.S.,
L
OW
-F
REQUENCY
F
ILTER
ON
Range
0.1 V Range
1.0 V – 300.0 V Ranges : (V
IN
x Freq.
≤
1.5x10
7
)
20 Hz – 30 Hz
1.10 + 0.15
1.10 + 0.03
30 Hz – 50 Hz
0.55 + 0.15
0.55 + 0.03
50 Hz – 200 Hz
0.23 + 0.15
0.23 + 0.03
200 Hz – 10 kHz
0.08 + 0.15
0.05 + 0.03
10 kHz – 50 kHz
0.23 + 0.15
0.10 + 0.05
50 kHz – 100 kHz
0.70 + 0.25
0.40 + 0.15
100 kHz – 300 kHz
4.00 + 0.60
4.00 + 0.60