www.vtiinstruments.com
VM2710A Command Dictionary
47
IEEE
488.2
C
OMMON
C
OMMANDS
*CLS
Purpose
Clears all status and event registers
Type
IEEE 488.2 Common Command
Command Syntax
*CLS
Command Parameters
N/A
*RST Value
N/A
Query Syntax
N/A
Query Parameters
N/A
Query Response
N/A
Description
This command clears the Status Event Register, Operation Status Register and the
Questionable Data/Signal Register. It also clears the OPC flag and clears all queues
(except the output queue).
Examples
Command / Query
Response (Description)
*CLS
(Clears all status and event registers)
Related Commands
N/A