NOTE:
The CIP solution level is calculated based on the CIP’s
pump capacity to exhaust the 3L of CIP
solution.
4.7 PAE alarms
All the PAE alarms resolve automatically and do not require any intervention.
PAE COMM (80h)
PCB and power supply communication error
PAE OVP SHUT (100h)
Over voltage protection
PAE OLP SHUT (200h)
Overload protection
PAE OTP SHUT (400h)
Module power supply error, over potential
PAE FAN FAIL (800h)
Power supply fan failure
PAE SMPS (1000h)
Switch mode power supply
PAE HIGH TEMP (2000h)
Module power supply error, too high
temperature
PAE ACP DOWN (4000h)
Power supply interrupted
PAE AC IP FZ (8000h)
Module power supply error cut off mains
power
If a PAE alarm is persistent over a period of 1 week, please contact Voltea.
4.8 ZERO EC ERR. (40000h)
Conductivity reading is 5 5 µS/cm or less for more than 30 seconds, unit goes to IDLE at the end of the
cycle.
A.
Water flowing through the unit.
o
Ensure feed line is open.
o
Check the water connection and compare it with the P&ID.
o
Check the manometer connected in the pure line. The pressure indicated should
increase during the pure step.
o
If consistent flow through the unit is confirmed, please go to troubleshooting event B.