10 |
P a g e
H.
Low Flow P.Alrm
setpoint and the pure line flow restrictor.
Tools: Adjustable wrench, a hex key (to push the flow restrictor out of the manifold).
o
Navigate to the Setpoint screen and check the
Low Flo P.Alrm
setpoint. Compare it
with Voltea’s recommendation for your application, adjust if needed.
o
Close the water supply to the DiEntry
o
Flush the unit with the feed water supply closed using the automated function
(Flush
module)
for 5 seconds to depressurize the system (skip this step if no water was
introduced to the unit after step F).
o
Check the color of the pure line flow restrictor. Confirm that the flow restrictor is in
line with Voltea’s recommendation for your application, replace if needed.
▪
Remove the module manifold.
▪
Remove the pure line connection and check the flow restrictor color.
Module manifold