24 |
P a g e
D.
Check the Electrical Conductivity (EC) probe for damage.
Tools: Philips screwdriver, calibration manifold, calibration solution.
o
Disconnect the EC probe from the system
▪
Unscrew the probe using a screwdriver, remove the top part of the probe as
shown in the picture below.
o
Connect the removed part to the calibration manifold probe.
o
Fill the calibration manifold, preferably with a solution of known electrical
conductivity.
o
Open the electrical cabinet and connect to the unit using Docklight, for more
information on Docklight please refer to the Reseller Manual (Chapter 9.2 Connection
to laptop and utilizing Docklight).
o
Click the following commands:
o
After clicking the second command the unit should respond with the measured EC:
▪
If the measured EC is 0 please contact Voltea