72
j) Capacity
Range
Resolution
Accuracy
60.00 nF
0.01 nF
±(4 % + 6)
600.0 nF
0.1 nF
6.000 µF
0.001 µF
60.00 µF
0.01 µF
600.0 µF
0.1 µF
6.000 mF
0.001 mF
±10 %
60.00 mF
0.01 mF
±13 %
600 V overload protection
The accuracy values are valid when REL mode is enabled
k) Diode test
Test voltage
Resolution
approx. 3.0 V/DC
0.001 V
Overload protection: 600 V; Test voltage: 1.8 mA typ.
l) Frequency “Hz” (electronic)
Range
Resolution
Accuracy
10 Hz – 99.99 Hz*
0.01 Hz
±(0.1 % + 5)
999.9 Hz
0.1 Hz
9.999 kHz
0.001 kHz
99.99 kHz
0.01 kHz
999.9 kHz
0.1 kHz
40.00 MHz
0.01 MHz
600 V overload protection
Signal level (without DC component):
≤100 kHz: 200 mV – 30 Vrms
>100 kHz – <1 MHz: 600 mV – 30 Vrms
≥1 MHz – <10 MHz: 1 V – 30 Vrms
10 MHz – 40 MHz: 1.8 V – 30 Vrms
* The frequency measuring range starts from 10 Hz
The standard measuring range includes frequencies ≤10 kHz.