Eccellenza Touch
VKI Publication #100309-001
47
Removing the Water Tank
1.
With the water tank drained and the service cord disconnected, remove the thermal
tape from the valves, and disconnect the power wires from each of the valves.
2.
Disconnect the outlet hoses from each of the outlet valves.
3.
Disconnect the inlet hose (coming from the inlet valve).
4.
Pull the tank towards you and partially out of the machine - you need to get access to
the tank top.
5.
Disconnect the wiring from the heating element.
6.
Disconnect the wiring from the three level probes.
7.
Disconnect the inline connector to the temperature probe.
8.
Pull the water tank completely out of the machine.
You will need to remove the overflow hose from the upper and lower
openings in the metal through which the hose is routed.
Replacing the Temperature Probe and Water Level Probes
1.
With the water tank removed, pull the temperature probe and/or the water level
probes out through the top of the tank lid. They are simply clipped in place.
2.
Slide the new temperature probe and/or water level probes completely into the tank
lid until they are clipped and locked in position.