V1410 / V1412 User Manual
94
Sidestream CO
2
Measurement Method
1
Connect the CO
2
module connector to the V1410 / V1412 CO2 connector and connect the sample
line to the sampling port on the patient breathing wye piece.
2
Push the CO
2
module connector into the CO
2
port until
you hear a click.
3
Respiration data can be achieved via ECG connections or EtCO2 measurements. When the
EtCO2 sensor is connected to the monitor, the measurement is prioritized to pull the respiration
data from the EtCO2 sensor.
4
EtCO2 and FiCO2 will be displayed on the V1410 / V1412 screen 15 seconds after a successful
detection of exhaled breath.
5
To activate the EtCO2 trace, select Menu
EtCO2
TRACE
ON.