V1410 / V1412 User Manual
47
24. Default Settings
When you select DEFAULT SETTINGS, all setting values are changed to the values saved at the factory
without you having to change the setting values for each parameter. The default settings are as follows:
Parameter
Item Name
VET_Large
VET_Small
ECG
LEAD
1st CH
LEAD II
LEAD II
GAIN
1st CH
10 mm/mV
10 mm/mV
SPEED
1st CH
25 mm/sec
25 mm/sec
CONFIG
CABLE
3 LEAD
3 LEAD
NOTCH
START
START
AREA
60Hz
60Hz
FILTER
ANIMAL
ANIMAL
ST
J + 80ms
J + 80ms
ARRHYTHMIA
STOP
STOP
PACEMAKER
STOP
STOP
HR CALC.
4
4
DEFIB. SYNC
OFF
OFF
ALARM
HIGH Limit
140 bpm
180 bpm
LOW Limit
70 bpm
70 bpm
TYPE
HIGHEST
HIGHEST
ST LEVEL
ALARM
HIGH
+5.0
+5.0
LOW
-5.0
-5.0
TYPE
HIGHEST
HIGHEST
SpO2
GAIN
X1
X1
SPEED
25 mm/sec
25 mm/sec
APG
APG OFF
APG OFF
TRACE
TRACE ON
TRACE ON
ALARM
HIGH Limit
OFF
OFF
LOW Limit
90 %
90 %
TYPE
HIGHEST
HIGHEST