30. Arrhythmia Display
The arrhythmia display can identify a problem with a patient’s ventricle.
NOTE
Although studied widely in human medicine, arrhythmia measurements have not been
validated in veterinary medicine and are offered only as an ancillary tool.
How to Analyze Arrhythmia
1
To analyze the arrhythmia, press the Menu button once,
select ECG
CONFIG
ARRHYTHMIA, and then select
START.
2
3
When arrhythmia is detected, an alarm sounds and a message is displayed in the arrhythmia
display window.
4
To cancel the audible arrhythmia alarm, press the EXIT button once. However, the message will
display continuously on the arrhythmia screen.
CAUTION
Because there are a great variety of species, and rhythms do vary between species,
the V1410 / V1412 arrhythmia analysis is only intended to be a helpful tool for which a
trained, qualified veterinary professional should ultimately decipher and prescribe
intervention based on their medical opinion.
The V1410 / V1412 may not always be able to detect arrhythmia in all patients. The
arrhythmia analysis is intended to be used for monitoring the patient’s heart rhythm, but
hospital staff must carefully monitor the patient’s condition for emergencies.
PVC detection screen at the start of arrhythmia analysis: