SpO2 Measurement Display
No.
Description
No.
Description
1
SpO2 Waveform Speed, Size
4
SpO2 Pulse Bar
2
SpO2 Waveform
5
SpO2 Measurement Numeric
3
SpO2 Alarm Range
How to Measure SpO2
1
Connect the SpO2 connector to the equipment and
attach the SpO2 sensor to the patient’s tongue, toe
webbing, lip, prepuce, vulva, or ear.
2
For stable measurement, minimize patient
movement.
NOTE
If measured with NIBP at the same time, the SpO2 probe should be positioned on the opposite limb.
Concurrent use of other medical devices affecting blood flow should be avoided. The probe should
be attached at a site that avoids those affected by medical interference.
●